Operation – Leica Biosystems ST5020 User Manual

Page 41

Advertising
background image

41

Leica ST5020

Time/Location

Event

Alarm ceases / switched off

Alarm type

The table below shows the specific events each individual alarm type is allocated to.

Instrument
being
switched on:

Drawers:

Staining:

Ovens:

Malfunctions
requiring
service:

Coverslipper:

Alarm 1

Alarm 3

Alarm 3

Alarm 3

Alarm 3

Alarm 1

Alarm 1

Alarm 1

Alarm 3

Alarm 3

Alarm 3

Alarm 2

Alarm 3

Alarm 3

All unload containers occupied when instrument is
switched on.

Switching on after power failure while staining
procedure was in progress.

Load/unload drawer open - time limit exceeded.

Load/unload drawer opened while gripper located
in drawer area.

Unload drawer is full or rack was placed into the
unload drawer.

Rack bearing a clip of a color not allocated to any
program placed into load station.

Rack bearing a white clip placed into load station.

Rack bearing an unknown clip placed into load
station.

Pause button has been pressed.

Unload drawer occupied, gripper attempting to
unload completed rack.

A program the last step of which is not allocated to
the unload drawer has been completed. - Rack
must be removed.

One of the ovens does not reach set temperature.
If the error occurs again, there will be a second
alarm after 60 seconds.

Hardware failure, e.g. motion error of transfer arm.
If the error occurs again after the instrument has
been switched back on, the alarm will immediately
sound again.

If the coverslipper requires an intervention by the
user to function or continue functioning.

-

When all racks have been re-
moved.

-

When dialog box is acknowl-
edged.

-

After drawer has been closed.

-

After drawer has been closed.

-

After rack has been removed, dia-
log box has been acknowledged.

-

After rack has been removed or
program allocated.

-

After rack has been removed or
program allocated.

-

After rack has been removed or
color allocated to unknown clip.

-

When leaving pause mode.

-

After dialog box has been ak-
knowledged (‘successfully re-
moved’).

-

After dialog box has been ak-
knowledged (‘successfully re-
moved’).

-

After dialog box has been ak-
knowledged

-

After dialog box has been ak-
knowledged and instrument swit-
ched off.

-

After dialog box has been ack-
nowledged

5.

Operation

Advertising