Setting up the instrument – Leica Biosystems ST5020 User Manual

Page 22

Advertising
background image

22

Instructions for Use, V1.9, RevD - 12/2012

Cover panel (26) must be removed from the
right inner wall of the instrument, so that the
transfer arm can access the oven stations.

To do so, remove six screws (27) (Allen key,
size 2.5) and remove cover panel (26).

Store screws and cover panel in safe place
for further use.

Important!

Prior to taking any further installa-
tion steps, switch the instrument off
and unplug from mains.

Remove cover panel (28) located to the right
of the electrical connections at the rear wall
of the instrument.

To remove panel (28), only remove screws
(29).

Do not loosen the two screws (30)!

Once the oven is installed, attach cover
panel (part of oven kit delivery) (17

in Fig. 10

)

with four screws (29) in such a way that the
voltage selected is visible from the outside.

Fig. 12

Fig. 13

26

27

27

28

29

29

29

29

30

4.

Setting up the instrument