Setting up the instrument – Leica Biosystems ST5020 User Manual

Page 22

Advertising
background image

22

Instructions for Use, V1.9, RevD - 12/2012

•

Cover panel (26) must be removed from the
right inner wall of the instrument, so that the
transfer arm can access the oven stations.

•

To do so, remove six screws (27) (Allen key,
size 2.5) and remove cover panel (26).

•

Store screws and cover panel in safe place
for further use.

Important!

Prior to taking any further installa-
tion steps, switch the instrument off
and unplug from mains.

•

Remove cover panel (28) located to the right
of the electrical connections at the rear wall
of the instrument.

•

To remove panel (28), only remove screws
(29).

Do not loosen the two screws (30)!

•

Once the oven is installed, attach cover
panel (part of oven kit delivery) (17

in Fig. 10

)

with four screws (29) in such a way that the
voltage selected is visible from the outside.

Fig. 12

Fig. 13

26

27

27

28

29

29

29

29

30

4.

Setting up the instrument