Setting up the instrument – Leica Biosystems ST5020 User Manual

Page 20

Advertising
background image

20

Instructions for Use, V1.9, RevD - 12/2012

Connecting the waste water hose

Outlet connection piece (10) can be turned
to point into different directions, depending
on whether the drain water pipe is located to
the left or to the right of the instrument.

• To adjust the connection piece, loosen

screw (11) of hose clamp (12) sufficiently so
that the connection piece (10) can be turned.
Adjust the connection piece as desired (al-
ways with the opening pointing slightly
downwards) and retighten screw (11).

Turn screw (15) to the left (use screwdriver)
until hose clamp (14) opens wide enough to
be slid over waste water hose (13).

Fit waste water hose (13) onto outlet con-
nection piece (10) and position hose clamp
(14) as shown in

Fig. 9

.

Turn screw (15) to the right until hose (13) is
secured tightly on connection piece (10) by
hose clamp (12).

Insert waste water hose (13) into drain pipe
and secure in position.

Fig. 9

7

13

14

15

Important!
When installing the drain hose al-
ways make sure there is a sufficient
descending gradient from the outlet
(connection piece (10)) on the instru-
ment to the drain pipe!

4.

Setting up the instrument