2 defining the program steps, Defining the program steps, Operation – Leica Biosystems ST5020 User Manual

Page 61

Advertising
background image

61

Leica ST5020

Fig. 56

For a complete definition of a program step, the following parameters
have to be entered:

1. Station in which the step is supposed to be carried out.

2. Reagent corresponding to that step (must be allocated to the station).

3. Duration of the step.

4. Step duration must be defined as exact / non-exact.

5. Dips (yes/no).

6. Oven temperature (optional)

5.6.2 Defining the program steps

Press DETAILS (

Fig. 54

) to access the PROGRAMMING menu (

Fig. 56

).

If a program is created from scratch, the list of the individual program
steps is completely empty.

The number in the STEP column indicates the order in which the sta-
tions will be used.
Press INSERT ROW - a new program row (step) is created in the row
below and highlighted in blue/red.

The total number of steps (new
programs automatically start from
number “1“) is automatically in-
creased by 1.
The DURATION, EXACT and DIP
columns already contain standard
settings (2) which must be entered
depending on the program step.

Code and name of the program (1)
are located above the table con-
taining the program steps.

Program steps are displayed in the
order in which they are supposed
to be carried out.

Up to

40

steps can be defined for

each individual program.

1

2

5.

Operation