Cleaning and maintenance, 1 cleaning the instrument, Cleaning the instrument – Leica Biosystems ST5020 User Manual

Page 75

Advertising
background image

75

Leica ST5020

6.

Cleaning and maintenance

6.1 Cleaning the instrument

Prior to cleaning the instrument, always switch off mains and
unplug the power cord!
When handling cleaning detergents, follow the instructions of
the manufacturer and make sure all laboratory regulations in
force in your country are complied with.
Dispose of used reagents according to the laboratory regula-
tions in force in your country!
Immediately wipe away spilled solvents (reagents)! Solvent
resistancy of the lid surfaces is limited if exposed to solvents
for a lengthy period of time!
Do not use any of the following for cleaning the outside sur-
faces of the instrument: alcohol, detergents containing alcohol
(window cleaner!), abrasive cleaning powders, solvents con-
taining acetone or xylene!
No liquid may come into contact with the electrical connec-
tions or spill into the interior of the instrument and/or the trans-
port arm!

Interior

Remove the reagent and wash vessels. Clean the stainless-steel inner
walls with a regular cleaner and rinse with water.

Transfer arm

The transfer arm contains sensitive electronic components. Therefore do
not clean with any liquid but simply wipe down with a slightly moistened
cloth.

Outside surfaces

Clean the painted outside surfaces with a mild detergent and subse-
quently wipe down with a moistened cloth.
Do not use any solvents for cleaning the outside surfaces and/or the lid!

TFT touch screen

Use a commercial screen cleaner for cleaning the touch screen. Use as
directed by the manufacturer.

Advertising