Different sound signals, in several intensity levels, can be allocated to
each alarm type.

ALARM TYPE X

→ SOUND Y

Allocates one of four different sounds to the alarm type. Each time the
button is pressed, a different sound is allocated.

SOUND INTENSITY = Z

Selects the intensity of the sound. For alarm type 1 settings from 0 to 4
are possible. Intensity level = 0: alarm sound switched off.
For alarm types 2 and 3 only intensity levels 3 and 4 can be set.
The level currently selected is indicated on the button.

TEST

The alarm type set (sound, intensity) will sound for approx. 10 secs.

5.4.2 Alarm

There are three different types of alarms, each one allocated to certain events. Alarm type
1 can be simply switched off, alarm types 2 and 3 cannot.

Alarm type 1:
Alarm 1 is generated when user intervention is
required.

Alarm type 2:
An error has occurred or user intervention is ur-
gently required.

Alarm type 3:
A serious error has occurred.
Follow the instructions provided by the error
message.
If the same error occurs repeatedly, call Leica
Technical Service.