2 alarm, Operation – Leica Biosystems ST5020 User Manual

Page 40

Advertising
background image

40

Instructions for Use, V1.9, RevD - 12/2012

Fig. 36

Different sound signals, in several intensity levels, can be allocated to
each alarm type.

ALARM TYPE X

→

→

→

→

→ SOUND Y

•

Allocates one of four different sounds to the alarm type. Each time the
button is pressed, a different sound is allocated.

SOUND INTENSITY = Z

•

Selects the intensity of the sound. For alarm type 1 settings from 0 to 4
are possible. Intensity level = 0: alarm sound switched off.
For alarm types 2 and 3 only intensity levels 3 and 4 can be set.
The level currently selected is indicated on the button.

TEST

•

The alarm type set (sound, intensity) will sound for approx. 10 secs.

5.4.2 Alarm

There are three different types of alarms, each one allocated to certain events. Alarm type
1 can be simply switched off, alarm types 2 and 3 cannot.

→

→

→

Alarm type 1:
Alarm 1 is generated when user intervention is
required.

Alarm type 2:
An error has occurred or user intervention is ur-
gently required.

Alarm type 3:
A serious error has occurred.
Follow the instructions provided by the error
message.
If the same error occurs repeatedly, call Leica
Technical Service.