Operation – Leica Biosystems ST5020 User Manual

Page 57

Advertising
background image

57

Leica ST5020

Indications and buttons corresponding to the different station types

Reagent station
Additional buttons

To open the reagent list (

chapter 5.4.7

) for allocating reagents to stations.

Press UPDATE REAGENT after a reagent has been changed, in order to

set the date of “Last change“ to the current date;

set the date in the “Use by date“ row according to the RMS;

Set the rack counter to “00“.

Additional indication

Reagent name:

Shows the reagent allocated.

If the RMS is on, the following parameters are displayed:

Date last change:

Date when the reagent was last changed.

Use by date:

Date when next change is required.

Rack counter:

Number of racks processed so far.

Racks max.:

Number of racks as per RMS that can still be
processed before the reagent needs to be
changed .

Even when the RMS is switched off, an internal log is kept in
the instrument. Therefore, all data displayed after switching the
instrument on are always up-to-date.

Oven station

Additional button

Pressing TEMPERATURE opens up the numerical keypad for entering an
oven temperature. Temperature settings between 40 and 70 °C (corre-
sponding to 104-178 °F) may be selected.

Additional indication

Heating off/on - indication controlled by staining program.
Temperature in degrees Centigrade/Fahrenheit.

Ovens 1 and 2 (stations 37/38) and ovens 3 and 4 (stations 39/40)
are linked. All settings made for one of the ovens will automati-
cally be applied to the second one, too.

Water station

Additional buttons:

none

Additional indication.

Water flow yes/no - indication controlled by staining program.

5.

Operation

Advertising