Safety – Leica Biosystems ST5020 User Manual

Page 8

Advertising
background image

8

Instructions for Use, V1.9, RevD - 12/2012

Warnings - Markings on the instrument itself

Markings on the instrument showing the warning triangle indicate that the correct
Instructions for Use (as defined in this manual) must be followed when operating or re-
placing the item marked.
Failure to adhere to these instructions may result in an accident, personal injury, damage
to the instrument or accessory equipment.

Some instrument surfaces, which become hot during operation are marked with
this warning label. Touching these surfaces may cause burns.

Warnings -handling reagents

Be careful when handling solvents!
Always wear rubber gloves and safety goggles when handling the chemicals used in this
instrument.
Reagents used for tissue infiltration can be both toxic and/or flammable.
Dispose of waste solvents with care according to local regulations and the waste man-
agement policy of the company or institution.

Warnings - Transport and Installation

The instrument may only be transported in an upright position.
Four people are needed to lift / carry the instrument.
Install the instrument on an even laboratory bench which must be absolutely level.
Do not expose the instrument to direct sunlight (windows).
The instrument MUST be connected to an earthed mains power outlet socket. The instru-
ment must not be connected to an extension cord without protective earth conductor.
The instrument will automatically adjust to the required voltage and frequency at the
place of installation.
If new ovens are installed, they must be adjusted to the voltage and frequency required at
the installation site.
The instrument must be set up in a well ventilated area, free from any ignition sources.
The chemicals to be used in the Leica ST5020 are both flammable and noxious. Do not op-
erate the instrument in rooms with explosion hazard.
If there is a significant difference in temperature between the warehousing and the in-
stallation site of the instrument and if at the same time there is a high air humidity level,
condensation water may form. In this case, a waiting period of at least two hours must be
observed before the instrument is switched on. Failure to adhere to this waiting period
may result in damage to the instrument.

2.

Safety

Advertising