Leica ST5020

Important!
If only one oven is installed, it must
be located on the right side (seen
from the rear of the instrument).

Place oven on rails (32) and insert.

Keep pushing until the connection pieces of
the oven (positioning pin (24), connector
strip (25), see

Fig. 11

) lock into the corre-

sponding connections (31) in the instrument.

Installing the oven

(continued)

Plug the instrument back into mains and
switch on. The oven modules are detected
automatically and are displayed accordingly
in ‘instrument status’.

Go to ‘stations options’ and set the desired
oven temperature.
(See

chapter 5.5.3, page 56

for further de-

tails).

Next, use the two screws (18) and washers
(19) to secure the oven (15) in the bores pro-
vided in the instrument for that purpose.

If a second oven has been ordered, install it
the same way as the first.

Fasten cover (item 17, Fig. 10) so that the
voltage range selected for the oven (15) is
the same as the engraved voltage range.

Fig. 14

Fig. 15

15

31

32

18, 19

Indication with
ovens installed

Indication prior to

installing the ovens

INSTRUMENT STATUS

Fig. 16

4.

Setting up the instrument