6 staining programs, Staining programs, Operation – Leica Biosystems ST5020 User Manual

Page 59

Advertising
background image

59

Leica ST5020

5.6 Staining programs

Up to 50 different staining
programs can be entered
and stored.

Existing programs may be
copied and/or edited.

The DURATION column dis-
plays the total duration of all
program steps (in h:min:sec).

The READY TO START col-
umn (yes/no) indicates
whether a program is ready
to be started.

Fig. 55

The PROGRAMS menu includes a table listing all staining programs stored.
In supervisor mode (

Fig. 54

) the column headers CLIP COLOR, ABBR. and PROGRAM NAME

are displayed as buttons. In addition, the buttons DETAILS, COPY, DELETE and NEW are dis-
played, i.e. programs can be edited in this screen.

Individual button functions

NEW creates a new (empty) program, located at the end of the pro-
gram list.

Programs selected to be copied are always copied to the last row of
the program list. All settings of copied programs are identical to those
of the original ones and can be edited. The clip color allocated is al-
ways WHITE.

Switches to the PROGRAMMING menu (

Fig. 57

) where the individual

program steps are displayed. In supervisor mode new program steps
may be edited.

5.

Operation

Advertising