6 inserting the activated carbon filter, 7 leveling the instrument, Inserting the activated carbon filter – Leica Biosystems ST5020 User Manual

Page 25: Leveling the instrument, Setting up the instrument

Advertising
background image

25

Leica ST5020

Once the activated carbon fil-
ter is inserted, push lever (35)
downwards to properly se-
cure the filter in the filter
housing.

4.6 Inserting the activated carbon filter

Make sure to insert the filter
with the correct side pointing
upwards: the three arrows on
the label (51) located at the
right of the filter must point to
front

.

Important for trouble-free
operation:
The sealing profiles (33) must
fit against the rear wall of the
filter housing (34).

Fig. 19

33

34

35

Fig. 18

Filter - inserted

51

4.

Setting up the instrument

4.7 Leveling the instrument

•

Once all accessories are installed (don’t forget the rear panel!), move
the instrument to its final position.

•

Prior to leveling the instrument, check whether the bench surface of
the laboratory bench where the instrument will be installed is flat and
exactly horizontal.

•

Otherwise, use the box level (standard delivery) to determine the high-
est point of the bench. Next, the instrument foot located at the highest
point of the bench must be screwed in almost completely (

see Fig. 21

).

- Then level the instrument adjusting each foot as needed.