44

Instructions for Use, V1.9, RevD - 12/2012

5.4.5 Movement

For

NUMBER OF DIPS

DIP AMPLITUDE

LIFTING SPEED

only the settings indicated on
the buttons can be selected.
The currently selected setting
shows on the button next to the
“=“ symbol. Press the button to
go to the next higher setting.

Any changes in this menu require administrator mode. For this reason, changes can only
be performed if no staining programs are active.
The parameters set in MOVEMENT are valid for all programs.

NUMBER OF DIPS:

Indicates how often the gripper dips the rack
into a reagent container before the rack is
immersed in that container. ‘Yes’ must be
selected in the DIPS column of the corre-
sponding program.

No dips are performed if the setting
“Number of Dips = ’0’“ is used, even
if ‘Yes’ is entered in the DIP column
in the program.

LIFT SPEED:

Indicates at what speed racks are removed
from reagent containers. The smaller the
numbers selected, the slower the speed. At
slower speeds, less reagent is carried over.

Fig. 40

5.

Operation

DIP AMPLITUDE:

Defines the stroke length of the arm when
dipping the rack (‘2’ = longer stroke than ‘1’).