Completing the program

•

Once the parameters EXACT and DIP have been selected, the defini-
tion of your step is complete.

•

Press INSERT ROW to create a new row for the next program step and
once again define all step parameters as described above.

•

Keep creating program steps until your program is complete.

•

Program steps whose definition is not correct (duplicate station
assignment, incorrect reagent,...) are highlighted in red in the
REAGENT column. Such steps must be corrected accordingly.

•

A program will not be ready to start until all details have been defined.

The last program step

Three different types of “last steps“ can be defined:

1. Last step = unload station.

Go to CHOOSE STATION and select EXIT as last station.

2. Last step = transfer station to Cover Slipper (optional).

go to CHOOSE STATION and select CV as last station.

The REAGENT column from the reagent list now allows selecting
between CV program 1 to CV program 4.
The transferred rack is automatically processed with the selected pro-
gram (1 ... 4) in the CV5030.

3. Last step = any station other than EXIT or CV.

Programs do not necessarily have to finish in the unload station, as a
particular staining protocol might e.g. have to end with a specific re-
agent not contained in the last station.

Be sure to define the DURATION of the last step. Otherwise it
will not be carried out (default duration 00:00 will be assigned).
Otherwise, the program cannot be started.