Setting up the instrument – Leica Biosystems ST5020 User Manual

Page 19

Advertising
background image

19

Leica ST5020

Fig. 8

Instrument rear panel

water in-/outlets

4.

Setting up the instrument

10

11

7

12

Fig. 7

8

9

10

Instrument rear panel

- tap water inlets

Screw tap water inlet hose (7 in

Fig. 8

) for

tap water stations 8 - 12 onto water tap and
insert the other end into inlet (8 in

Fig. 7

) .

If station no. 7 (

Fig. 3

) will also be used as a

tap water station, connect a second tap
water inlet hose (optional accessory) to inlet
(9).

Inlet (9) only provides water (or any
other medium supplied via a sepa-
rate connection) to station no. 7. To
use station 7 as a water station, a
wash vessel must be inserted and
station 7 must be defined as a ‘tap
water’ or ‘fully-de-ionized water’
station (DI water station) in ‘stations
options’. (see also

Chapter 5.5.3

)