Operation – Leica Biosystems ST5020 User Manual

Page 48

Advertising
background image

48

Instructions for Use, V1.9, RevD - 12/2012

The RMS log

→

→

The Leica ST5020 is equipped with a

R

eagent

M

anagement System (RMS) graphically

representing processing history of reagents in ‘Instrument Status’ (

see chapter 5.5.1

).

These reagent data are stored in the RMS log table.

The RMS log table lists all re-
agent stations, the reagents al-
located to each station as well
as their processing history.

Fig. 45

Information contained in a table row:

Station
Shows the numbers of those stations defined as
reagent stations.

Reagent
Reagent currently in a station or to be added to
that station. The Leica ST5020 does not recog-
nize reagents automatically - reagents are allo-
cated by entry into the reagent list.

Last change
Date when the reagent was last
changed.

Racks
Number of racks that have been
processed in a particular sta-
tion since the reagent was last
changed.

Racks max.
Maximum number of racks that should be pro-
cessed in a particular station without the re-
agent being changed.

Days
Number of days since the reagent was last
changed.

Days max.
Maximum number of days the reagent should
remain in the station without being changed.

→

→