Design and Function

2.5 Electrical equipment

Operating voltage: See nameplate (page 16), 50/60 Hz

Current consumption: See nameplate (page 17 )

Protection class 1, i.e. operating insulation with PE conductor in accordance with EN

61010

Protection type IP 20 acc. to EN 60 529

Interference-suppressed acc. to EN 55011 class B

Appliance fuse: Fusible link 250 V/15 A quick-blow

The temperature controller is protected with a miniature fuse 100 mA (200 mA at 115 V)

The CO

2

controller is equipped with a 6.3 A miniature fuse.

2.6 Connections

1

3

4

5

6

2

CO

2

Fig. 5

Connections on rear of appliance

1 Mains lead

Fig. 5

Fig. 5

Connections on rear of appliance

Connections on rear of appliance

2 CO

2

connection (only for models with CO

2

supply)

3 USB connection (optional

22

connection (only for models with CO

connection (only for models with CO

ethernet connection, see page

connection (only for models with CO

connection (only for models with CO

22

supply)

supply)

57 )

57

57

4 Printer connection (see page 51 )

5 Water connection (see page 23

Water connection (see page

Water connection (see page )

23

23

6 Fresh air connection (only for models with CO

Water connection (see page

Water connection (see page

2

supply)

2.6.1 Electrical connection

Observe the country-specific regulations when connecting (e. g. in Germany DIN VDE 0100

with residual current circuit breaker).
This appliance is intended for operation on an electrical power system with a system imped-

ance Z

max

of a maximum of 0.292 ohm at the point of transfer (service line). The operator

must ensure that the cooled incubator is operated only on an electrical power system that

meets these requirements. If necessary, you can ask your local energy supply company what

the system impedance is.