4 adjusting door, 5 repairs and service – Memmert ICH 256 Climate Chambers (Generation 2003) User Manual

Page 68

Advertising
background image

68

Maintenance and Servicing

8.4 Adjusting door

A well-closing door is indispensable for cooled

incubators. Thanks to the combination of a

chamber seal and a door seal, MEMMERT

appliances optimally guarantee the tight closing

of the door. In permanent operation, the flexible

seal material might begin to sag. To ensure that

the door closes exactly despite this, an adjust-

ment may be necessary ( Fig. 39 ).
The top section (1) of the door hinge can be

moved slightly in the direction of the arrow after

the two screws (2) at the top and bottom of the

door have been undone. Afterwards, tighten the

screws again.
Adjusting door:
1. Undo the headless screw (5) (with a jolting

motion, as it is fixed with locking paint).

2. Adjust the door by turning the eccentric (3)

with a screwdriver.

3. Apply the locking paint to the headless

screw and it back tight.

The locking plate ( Fig. 40 ) can also be adjusted

in the direction of the arrows after undoing the

screw (6).
1. Undo screw (6).
2. Move locking plate (7) in direction of arrow.
3. Tighten the screws again.

8.5 Repairs and Service

Warning!

After removing covers,

voltage-carrying parts may

be exposed. You may receive

an electric shock if you touch

these parts. Disconnect the

mains plug before removing

any covers. Any work

inside the unit may only

be performed by qualified

electricians.
Repairs and service work are

described in a separate service

manual for the ICH cooled incubator.

2

1

3

4

5

Fig. 39

Adjusting door

1 Upper section of the door hinge

Fig. 39

Fig. 39

Adjusting door

Adjusting door

2 Mounting bolts

Upper section of the door hinge

Upper section of the door hinge

3 Eccentric tappet

4 Eccentric socket

Eccentric tappet

Eccentric tappet

5 Headless screw

7

6

Adjusting locking plate

6 Screw

7 Locking plate

Advertising