Advanced Functions

Due to this automatic defrosting, there is a minor brief increase in the chamber temperature

at regular intervals. If you would like to further reduce this detraction in performance, you can

decrease the defrosting frequency, e.g. to every 24 hours.
In this case, please observe if there is a permanent drop in the cooling performance or a

strong fluctuation of the actual value, which may be an indication that the cooling unit is icing

over. If this is the case, please set the automatic defrosting system one level higher.
If humidity/room temperature are particularly high, it is possible that the factory setting

for defrosting, 12 hours, is not sufficient. If this is the case, you should set a more frequent

defrosting interval, e.g. every 6 hours.
Automatic defrosting is disabled with the parameter OFF. When operating at low tempera-

tures, this causes the cooling unit to ice over over time. Regular defrosting needs to be carried

out in order to prevent damage to the cooling system.

7.7 User ID card (optionally available as an accessory)

The device number of the appliance and a

unique user number are stored in

encrypted form on the user ID card ( Fig.

27 ). The user ID card therefore works

only in the appliance with the

corresponding serial number.
To use it, insert the card into the chip

card reader (see page 11 ) on the

appliance control panel.
If the user ID card has been inserted, the

additional menu item ID-LOCK appears

in the SETUP menu. If this setting is

changed to ON, the appliance is locked

against all adjustments once the chip

card is removed.

Locking with the user ID card is displayed with the lit up icon

on the control

panel (see also page 28 ).

Every login process with the user ID card is logged in the internal flash memory of the

controller.

7.8 Calibration

7.8.1

Temperature calibration

The cooled incubator can be calibrated customer-specifically using three calibration

temperatures of your choice:

CAL.1 Temperature calibration at low temperature (e.g. 5 °C)

CAL.2 Temperature calibration at medium temperature (e.g. 25 °C)

CAL.3 Temperature calibration at high temperature (e.g. 50 °C)

access

authority

card

Name:

ID:

_____________________

_____________________

Fig. 27

User ID Card