Advanced Functions

Example: Humidity deviation at 80 % should be corrected.
1. Set humidity balance point in the SETUP to RH 80 and set the accompanying compensation

correction value to 0.0 %rh:

PRINT

SETUP

loop

t3

t4

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

3

4

2

1

STERI

DEFRO

°C

°C

rh

mb

%

CO

mb

2

IN 1

IN 2

OUT

MIN

AUTO

MAX

2. With a calibrated reference instrument, an actual humidity of 78 % rh is measured at

normal operation, with a defined setpoint humidity of 80 % rh.

3. Set the compensation correction value in the SETUP for RH 80 to –2.0 % rh:

PRINT

SETUP

loop

t3

t4

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

3

4

2

1

STERI

DEFRO

°C

°C

rh

mb

%

CO

mb

2

IN 2

OUT

IN 1

IN 2

OUT

MIN

AUTO

MAX

4. After the calibration procedure, the reference instrument should display 80.0 %.
With RH 20 a further comparison can be programmed at 20 % relative humidity.

If all compensation correction values are set to 0.0 % rh, the factory calibration settings

are restored.

7.8.3 CO

2

calibration

Customer calibration of appliance on controller by means of three CO

2

points:

CO2 5

CO

2

calibration at 5 % CO

2

content

CO2 10

CO

2

calibration at 10 % CO

2

content

CO2 15

CO

2

calibration at 15 % CO

2

content

For each selected balance point, a positive or negative compensation correction value can be

set.

co2 5

-0,4 %

20%

0%

10%

co2 10

+1,5 %

co2 15

-0,7 %

Default calibration

Fig. 30

CO

2

calibration (example)