Design and Function

D

A

E

B

F

C

Fig. 7

Dimensions of ICH cooled incubators

2.11 Ambient conditions

The cooled incubator may only be used in enclosed rooms and under the following ambient

conditions:

Ambient temperature: 5 ºC to 28 ºC

Humidity:

max. 80 %

Degree of pollution:

2

Altitude of installation: max. 2000 m above sea level

The cooled incubator may not be used in areas where there is a risk of explosions. The

ambient air must not contain any explosive dusts, gases, vapours or gas-air mixtures. The

cooled incubator is not explosion-proof.

Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to

sedimentation in the interior and, as a consequence, could result in short circuits or dam-

age to electrical parts. For this reason, sufficient measures are to be taken to prevent large

clouds of dust or aggressive vapours from developing.

2.12 Accessories

One sliding shelf

Water supply tank and connection tube

Additional for cooled incubators with CO

22

supply:

supply:

Gas pressure tube