6 close statements for ramp segments – Memmert ICH 256 Climate Chambers (Generation 2003) User Manual

Page 39

Advertising
background image

39

Operation and Control

5.10.6 Close statements for ramp segments

Each ramp must be completed with a close statement connecting the ramp to the next one.

These commands thus control the programme sequence:

NEXT

Connect the next programme segment.
SET-POINT WAIT (T – temperature)

Wait until the setpoint temperature has been reached.

The appliance starts the next programme segment only when the

programmed set temperature has been reached, even if the set

heating up time has already elapsed.
SET-POINT WAIT (H – humidity)

Wait until setpoint humidity has been reached.

Appliance starts the next programme segment only when the

programmed setpoint humidity has been reached, even if the set

time has already elapsed.
SET-POINT WAIT (TH – temperature and humidity)

SET-POINT WAIT (TH – temperature and humidity

SET-POINT WAIT (TH – temperature and humidity

Wait until setpoint temperature and setpoint humidity have been

reached. The appliance starts the next programme segment only

when the programmed setpoint temperature and programmed

setpoint humidity have been reached, even if the set time has

already elapsed.
Ramp repeat function

The programme entered is repeated after it has run through all

programmed segments.

1-99 = repetitions

loop = endless repeat function
Programme end while maintaining the temperature and humidity

of last programme ramp
Programme end, switching off the heating / cooling function and

humidification

Close

command

ramp

segment

No. 5

end

Segment5

°C

t=time

Delayed

programme start Segment1

Close

command

ramp

segment

No. 1

spwt (t)

Segment2

Close

command

ramp

segment

No. 2

next

Segment3

Close

command

ramp

segment

No. 3

spwt (tH)

Segment4

Close

command

ramp

segment

No. 4

next

Fig. 19

Schematic example of the use of ramp segment close statements