Maintenance and Servicing

8. Maintenance and Servicing

8.1 Cleaning

Warning!

Danger of injury by electric shock. Before any cleaning work, pull

out the mains plug.

8.1.1 Interior and metal surfaces

Regular cleaning of the easy-to-clean interior prevents

build-up of material remains that could impair the

appearance and functionality of the stainless steel

chamber over time.
To clean the interior, the fan cover can be removed by

pulling it slightly forwards after the two fixing screws

have been screwed out ( Fig. 31 ). The metal surfaces of

the chamber can be cleaned with normal stainless steel

cleaning agents.
Make sure that no rusty objects come into contact with

the working chamber or with the stainless steel housing.

Rust deposits can lead to an infection of the stainless

steel. If rust spots should appear on the surface of the

working chamber due to impurities, immediately clean

and polish the affected area.

8.1.2 Plastic parts

Do not clean the operating panel, the plastic input module and other plastic parts of the ap-

pliance with caustic or solvent-based cleaning agents.

8.1.3 Peltier module

In order to guarantee perfect function and long lifetime

of the Peltier dehumidification modules on the rear side

of the appliance, it is absolutely essential to remove dust

deposits from the heat sink (with a vacuum cleaner,

paintbrush or bottle brush, depending on the amount).
To make cleaning easier, the protective cover can be

removed after the screws have been loosened ( Fig. 32 ).

Fig. 31

To take off the fan

cover, remove two screws

Fig. 32

Cover of the Peltier

cooling modules

Fig. 32

Fig. 32

Advertising