Memmert ICH 256 Climate Chambers (Generation 2003) User Manual

Page 61

Advertising
background image

61

Advanced Functions

3. Set compensation correction value for CAL.2 under SETUP to - 0.4 °C:

h

°C

°C

4. After the calibration procedure, the reference instrument should display 35 °C.
5. With CAL.1, another compensation temperature below CAL.2 can be programmed in the

same way, and with CAL.3, one above this.

If all compensation correction values are set to 0.0 °C, the factory calibration settings are

restored.

7.8.2 Humidity calibration

The cooled incubator can be calibrated for the individual customer by means of two balance

points at 20 and at 80 % relative humidity. For each selected balance point, a positive or nega-

tive compensation correction value can be set between –5 % and +5 %.

rh 20

+3,0 %

rh 80

-2,0 %

60%

40%

0%

20%

80%

100%

Default calibrati

on

Fig. 29

Humidity calibration (example)

Setting:

Setting:

1. Set the desired humidity balance point in the SETUP (see page 53 ) and set the accompany-

ing compensation correction value to 0.0°C.

2. With a reference instrument, measure the deviation in the stationary state in the selected

humidity balance point.

3. Under SETUP, adjust the compensation correction value. If the measured reference humid-

ity is too low, the compensation correction value must be set with a negative sign.

4. Perform a control measurement with the reference instrument.
The procedure can be performed with humidity balance points of 20 % rh and 80 % rh.