Operation and Control

14. Setting the fan speed of the second ramp segment:

Turn the push-turn control to the right until the fan

display flashes.

Hold down the SET key and set the desired fan speed of

50 % (five bars lit up) with the push-turn control.

15. Setting the relative humidity of the second ramp seg-

ment:

Turn the push-turn control to the right until the humid-

ity display flashes.

Hold down the SET key and set the desired humidity

setpoint to 70.0 % rh with the push-turn control.

rh

%

16. Setting the CO

2

content of the second ramp segment

(only for appliances with CO

2

supply, otherwise please

continue with point 17):

Turn the push-turn control to the right until the CO

2

display flashes.

Hold down the SET key and set the desired CO

2

setpoint

to 15.0% with the push-turn control.

CO

2

17. Set lighting of the second ramp segment (only for mod-

els with lighting, otherwise please continue with point

18):

Turn the push-turn control to the left until the lamp uv

display appears.

Hold down the SET key and choose on with the push-

turn control.

Turn the push-turn control to the left until the lamp dl

display appears.

Hold down the SET key and set to On with the push-turn

control.

18. Setting the close statement of the second ramp seg-

ment:

Turn the push-turn control to the right until a segment

close statement, e.g. end, appears.

Hold down the SET key and set the close statement next

with the push/turn control.

19. Setting the duration of the third ramp segment:

Select the time display with the push-turn control.

Hold down the SET key and set the time to 1.00 with the

push-turn control.

h

20. Setting the temperature of the third ramp segment:

Turn the push-turn control to the right until the tem-

perature display flashes.

Hold down the SET key and set to 20.0 °C with the push-

turn control.

°C