Warning Messages and Malfunctions

6.3 Power failure

In case of a power failure, the cooled incubator operates as follows:

In normal

and week time switch operating modes

and week time switch operating modes

After the power supply has been restored, operation is continued with the parameters set.

The time and duration of the power failure are documented in the log memory.

In Programming mode

In Programming mode

After a power failure of less than 60 minutes, the current programme is continued from

the point at which it was interrupted. The time and duration of the power failure are

documented in the log memory (see page 58 ).

In case of a power failure of more than 60 minutes, the cooled incubator starts in manual

operating mode for safety reasons and all setpoint values are changed to safe default

values (see table below).

For remote operation

For remote operation

If there is a power failure in remote operation, the appliance starts in the normal operating

mode for safety reasons and all setpoint values are changed to safe default values (see table).

The programme can only be continued from the computer. The time and duration of the

power failure are documented in the log memory.

Parameters

Default value

Temperature

20 °C

Fan speed

100 %

CO

2

0 %

rh

off