11 during operation – Memmert ICH 256 Climate Chambers (Generation 2003) User Manual

Page 45

Advertising
background image

45

Operation and Control

5.11 During operation

Regularly check the water level. If necessary, add distilled water.
Warning messages during operation: See page 47 .

CO

22

mode

In the heating up phase, the CO

2

controller is initially deactivated. The CO

2

inlet is interrupted

during this period. About 5 minutes after the setpoint temperature has been reached, the CO

2

control begins measuring and CO

2

gas is introduced into the chamber via a sterile filter. The

setpoint can be set from 0 to 20 % in 0.1 % steps.

Displays in CO

Displays in CO

22

mode:

displayed during the heating up phase of the CO

2

sensor. In the CO

2

display, CO2 is shown.

displayed during the zero balance adjustment.

CO

2

After the setpoint temperature has been reached,

the CO

2

concentration is displayed in % depending

on the setting.

displayed if the CO

2

concentration exceeds the

defined setpoint by at least 1 % for more than 3

minutes. If the concentration is higher, the CO

2

display and the icon flash.

icon flash.

In this case, you should open the door for 30 sec.

and wait to see if the controller steadily adjusts to

the setpoint. If the error occurs again, contact the

customer service.

This monitoring function only starts

working once the CO

2

setpoint has been

reached.

displayed if the gas bottle is empty. In this case,

connect a new gas bottles (see page 23).

The pressure in the gas bottles is a constant

approx. 57 bar at 20 °C ambient tempera-

ture. The pressure in the bottle cannot be

used to determine its filling level, since the

pressure only drops immediately before the

bottle is completely empty.