Warning Messages and Malfunctions

Error

Possible cause

Remedy

icon flashes

...

Temperature protection

(TWW, ASF) has

triggered

Increase temperature difference between

monitoring and working temperature

(see page 53 ).

Replace Pt100 temperature sensor of

monitoring controller, if necessary (see

service manual)

... and rh

empty

Water supply tank empty Set humidity setpoint to OFF, fill up distilled

water, then reset humidity setpoint back to

desired value.

... and rh

over

Humidity setpoint

exceeded

Open the door for 30 sec. and wait to see

if the controller then steadily adjusts to the

setpoint. If the error occurs again, contact

the customer service.

... and co2

over

CO

2

setpoint is exceeded Open the door for 30 sec. and wait to see if

the controller steadily adjusts to the setpoint.

If the error occurs again, contact the

customer service.

... and error

Autozero

Autozero pump

defect

CO

2

controller defect

Contact the customer service and refer to the

service manual.

Error display
(E...) in display

Appliance/system error

Contact the customer service and refer to the

service manual.

Error message

CONF
(is displayed

for only 10 sec.

after switching

on)

Checksum error (error

when saving setpoint

values)

The error can be rectified by the controller

itself after a setpoint parameter has again

been saved. If the error should continue

occurring or cannot be rectified: Contact the

customer service and replace the controller

(see service manual).

Chamber fan

does not work

Miniature fuse or mains

adapter faulty

Contact the customer service and refer to the

service manual.

Heating icon

does not light

up

Ambient temperature

too high

Set up appliance in a cooler room,

minimum setpoint temperature =

ambient temperature + 8 °C

Temperature in

appliance higher

than defined setpoint

temperature

Wait until the appliance has cooled down.

symbol is

symbol is

permanently on

Temperature protection

(TB) has triggered

Remedy: See page 54