Advanced Functions

Tolerance band ASF
Adjustment range: 0.5 to 5°C (see page 53 )

Language

Language
Setting: GERMAN, English, franCAIS, ESPANOL and italIANO

Corrective values
CAL 1-3, rh20, RH80, CO2 5, CO2 10, CO2 15
for customer-side calibration of temperature, humidity and

CO

2

(see chapter " Calibration " on page 59 )

Leave setup

Leave setup
Save all settings and leave the SETUP mode
The realtime clock which is set in the SETUP contains the date and clock time. It is used

for logging purposes in accordance with GLP. Date and clock time are specified on the log

printout. On graphical printouts, the time axis is labelled with the realtime. The clock is

battery-buffered and is independent of the mains connection. The integrated lithium battery

of the type CR 2032 has a lifetime of approx. 10 years.

7.3 Temperature monitoring

The monitoring temperature is measured via a separate Pt100 temperature sensor in the

chamber interior. The monitoring unit is used to protect the chamber load and as a protection

for the appliance and surroundings.
The appliance is equipped with a double overtemperature protection (mechanical/electronic)

in accordance with DIN 12880.

°C

MIN

AUTO

MAX

1

2

3

1

6
5

4

Fig. 22

Temperature monitoring display

1 Visual

Fig. 22

Fig. 22

alarm icon lit up: TB

Temperature monitoring display

Temperature monitoring display

alarm flashing: TWW

Temperature monitoring display

Temperature monitoring display

alarm, ASF alarm

2 Undertemperature protection

alarm icon lit up: TB

alarm icon lit up: TB alarm flashing: TWW

alarm flashing: TWW

3 Automatic temperature monitor (ASF, see page 56 )

56

56

4 Overtemperature protection (TWW, TWB, see page 54 )

54

54

5 Trigger temperature

6 Acoustic alarm icon

Trigger temperature

Trigger temperature