Operation and Control

7. Setting the fan speed of the first ramp segment:

Turn the push-turn control to the right until the fan

display flashes.

Hold down the SET key and set the desired fan speed of

30 % (three bars lit up) with the push-turn control.

set

off

on

push

card

4

3

2

1

mb

CO

mb

2

IN 1

IN 2

OUT

IN 1

IN 2

OUT

PRINT

SETUP

loop

t3

t4

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

°C

°C

MIN

AUTO

MAX

h

rh

DEFRO

STERI

%

1

2

3

4

5 6 7

8

9

10

11

13

14

15

16

17

8. Setting the relative humidity of the first ramp segment:

Turn the push-turn control to the right until the humidity

display flashes.

Hold down the SET key and set the desired humidity

setpoint to 70.0 % rh with the push-turn control.

%

rh

9. Setting the CO

2

content of the first ramp segment (only

for appliances with CO

2

supply, otherwise please con-

tinue with point 10):

Turn the push-turn control to the right until the CO

2

display flashes.

Hold down the SET key and set the desired CO

2

setpoint

to 5.0% with the push-turn control.

CO

2

10. Set lighting of the first ramp segment (only for models

with lighting, otherwise please continue with item11):

Turn the push-turn control to the left until the lamp dl

display appears.

Hold down the SET key and choose on with the push-

turn control.

11. Setting the close statement of the first ramp segment:

Turn the push/turn control to the right until a segment

close statement, e.g. end, appears.

Hold down the SET key and set the close statement

SPWT [TH] with the push/turn control.

12. Setting the duration of the second ramp segment:

Turn the push-turn control further to the right until the

time display flashes.

Hold down the SET key and set time to 0.45 with the

push-turn control.

h

13. Setting the temperature of the second ramp segment:

Turn the push-turn control to the right until the tem-

perature display flashes.

Hold down the SET key and set the desired temperature

setpoint to37.0°C with the push-turn control.

°C