Maintenance and Servicing

8.1.4 Cooling compressor

In order to guarantee perfect

function and long lifetime of the

refrigeration unit, it is absolutely

essential to remove dust deposits

from the condenser ( Fig. 33 , no. 1)

(with a vacuum cleaner, paintbrush

or bottle brush, depending on the

amount).
After the screws (2) have been

loosened, the air inlet cover on the

front (3) can be removed and the

condenser can be cleaned.
After the screws have been loos-

ened, the air inlet cover on the

rear (4) can be removed as well,

to clean the refrigeration module

from both sides if necessary.

8.2 Regular maintenance

Every three months

Every three months

Check that the door seals fit tightly and adjust the door if necessary (see page 68 ).

In permanent mode: Grease the moving parts of the doors (hinges and lock) with thin

silicon grease and check that the hinge screws are not loose.

Annually

Annually

Grease the moving parts of the doors (hinges and lock) with thin silicon grease and check

that the hinge screws are not loose.

Check that door seals fit tightly and adjust the door if necessary (see page 68 ).

Check safety labels (warning labels on the door) (see page 8 ). Replace them if they are

unrecognisable. They can be ordered from the Memmert customer service.

Check and replace the sterile filter if it is dirty (only models with CO

2

supply, see service

manual)

Additional maintenance every two years, for models with CO

Additional maintenance every two years, for models with CO

2

supply

supply

Replace all sterile filters (see service manual)

1

2

3

4

Cooling compressor