Advanced Functions

7.3.1 Mechanical temperature monitoring: Temperature limiter (TB)

The cooled incubator is equipped with a mechanical temperature limiter (TB) of protection

class 1 in accordance with DIN 12880.
If the electronic monitoring unit should fail during operation and the factory-set maximum

temperature is exceeded by approx. 20 °C, the temperature limiter, as the final protective

measure, switches off the heating permanently. As a warning, the

icon lights up.

icon lights up.

Error rectification after the TB has been triggered:

Error rectification after the TB has been triggered:

1. Switch off the appliance and leave to cool down.
2. Reset the TB by pressing the red button, located

on the rear side of the unit (next to the power

cord, see Fig. 23 ). Press until you hear a „click“

sound.

7.3.2 Electronic temperature monitoring

(TWW)

Temperature monitoring can be adjusted indepen-

dently from the operating modes.
The manually set monitoring temperature

MIN

and

and

MAX

the overtemperature control is monitored by an

adjustable over/undertemperature controller (TWW) protection class 3.3 acc. to DIN 12880.
If the manually set monitoring temperature

MAX

is exceeded, the TWW takes over

temperature control and begins to regulate the monitoring temperature ( Fig. 24 ).

The alarm icon flashes as a warning .

Setting MAX

Set
temperature

t

°C

Controller error

Emergency operation

Fig. 24

Schematic diagram of how the TWW temperature monitoring works

The monitoring temperature must always be set sufficiently high above the

maximum working temperature.

If the acoustic alarm is switched on in the SETUP, the TWW alarm is additionally signalled by

an intermittent tone. If the SET key is pressed, the acoustic alarm can be temporarily switched

off until the next alarm event occurs.

Fig. 23

TB Reset