Advanced Functions

7.5 Log memory

The controller continually logs all relevant measured values, settings and error messages at

1-minute intervals.
The internal log memory is listed as a ring buffer, i e. the oldest log data is always overwritten

automatically with new data.
The logging function cannot be switched off and is always active. The measured data are

stored in the controller, safe from manipulation. For documentation purposes, the controller

memory can be read out via the Celsius control software. Each dataset is stored with a unique

timestamp.
The internal log memory has a size of 1024 kB. This corresponds to a storage capacity of

about three months in permanent operation.
If the power supply is interrupted, the time of the power cut and the return of voltage are

stored in the controller.

Reading in the log memory to the computer via

Reading in the log memory to the computer via USB interface

The log data can be read out either via the USB interface and then printed out from the

computer or via a connected PCL3 compatible printer.
Using the “Celsius“ software, the log memory of the controller can be read out to a computer

and from there be displayed graphically, printed out and stored.

The log memory of the controller is not modified or deleted by reading it out.

Printing out log memory

Printing out log memory

(see also chapter " Printer " on page

Printer

Printer

51 )

For a printout, the GLP header is also printed automatically, containing the following details:

Date of printout

Period of log

Consecutive page numbers

Serial numbers and appliance name

If the printer is not ready (e. g. ink cartridge or paper tray empty), no log data will be

lost. Multiple printouts can also be made, since the log memory is not deleted after

printing.

7.6 Automatic defrosting system

The integrated automatic defrosting system for the cooling unit ensures perfect operation of

the cooled incubator at low temperatures and in permanent operation. The time values for

the automatic defrosting system can be set in the submenu

SETUP

under

under defrost (see page

defrost

defrost

52 ).
Due to long-term operation with a working temperature below +15 °C or due to a damp

chamber load and/or the door opening frequently, ice can form in the interior over time.

Heavy icing may impair the function of the appliance and could damage the cooling system.

In this case, the working chamber should be defrosted. This can be achieved by short heating

of the appliance (to 30-40 °C) or by switching it off for a longer period of time, e.g. over

night. The resulting melting water can best be collected with a cloth at the front edge of the

working chamber. Afterwards, the smooth surface of the interior can easily be cleaned.