Memmert ICH 256 Climate Chambers (Generation 2003) User Manual

Page 18

Advertising
background image

18

Design and Function

Technical data

without interior

lighting

with interior

lighting

max. load per

sliding shelve [kg]

30

max. load per appliance

[kg]

80

Temperature

Temperature recording is done by means of Pt100 in a 4-wire

circuit

Adjustment range normal operation: -10 °C to 60 °C, for mod-

els with CO

2

supply or interior lighting 0 °C to 60 °C

Adjustment precision: 0.1 °C

Deviation in time: max. ±0.1 °C at 37 °C

Spatial deviation: max. ±0,3 °C at 37 °C

Humidity

Relative humidity in the chamber is measured by a capacitive

humidity sensor and displayed digitally in percent. The meas-

urement precision of the humidity sensor is 1 % rh

Adjustment range: 10 to 80 % rh

Adjustment precision: 1 % rh

Display range: 10 to 98 % rh

Deviation in time: max. ±2 % rh

CO

2

The CO

2

content is determined

through an NDIR measuring pro-

cedure, controlled constantly by a

microprocessor and displayed digit-

ally in percent

Adjustment range: 0 to 20 %

Adjustment precision: 0.1 %

Deviation in time: max. ±0.1 %

Spatial deviation: max. ±0.3 %

Interior lighting

Daylight: LT 18W/965
UV: LT 18W/009 UV