3 safety labels, 4 requirements of the operating personnel – Memmert ICH 256 Climate Chambers (Generation 2003) User Manual

Page 8

Advertising
background image

8

Safety Regulations

1.2.2 Additional safety regulations for cooled incubators with CO

2

supply.

Warning!

Danger of suffocation. CO

2

can have a suffocating effect in high

concentrations. In normal operation, the cooled incubator gives

off small amounts of CO

2

to its surroundings. You should therefore

ensure that the room in which it is installed is properly ventilated.

Always close the stop valve or pressure reducer on the gas bottle if

there is no gas bottle connected or if the bottle connected is empty.
Warning!

High concentrations of CO

2

can cause cold burns or frostbite.

Avoid contact with CO

2

gas to the eyes and skin.

Warning!

Gas bottles may burst or explode at high temperature. Keep the gas

bottles away from open flames. Do not store gas bottles at or above

50 °C and ensure that the location is always well-ventilated. Prevent

water from penetrating as well as backflow into the gas bottles. It is

essential that you read the safety notes and regulations of the gas

suppliers.

CO

2

is not a dangerous substance in terms of the German Hazardous Substances Ordinance

(GefStoffV). You should nevertheless familiarise yourself with the applicable safety regulations

prior to handling such gas bottles.

1.3 Safety labels

Warning signs for dangerous UV radiation in the working

area are attached to the doors or cooled incubators

with interior UV lighting ( Fig. 1 ). They indicate that

either the lighting has to be switched off or UVsafety

goggles are required when the door is opened.
These stickers must not be removed and must always

be well visible. If they become unrecognisable or if they

peel off, they must be replaced. They can be ordered

from Memmert customer service.

1.4 Requirements of the operating

personnel

The cooled incubator may only be operated and maintained by persons who are of legal age,

and who have received relevant instructions. Personnel who are to be trained, instructed or

who are undergoing general training may only work with the appliance under the continuous

supervision of an experienced person.
The cooled incubator may only be transported (with a forklift truck or manual pallet jack)

by persons, who are trained for this work and are familiar with the corresponding safety

regulations.
Repairs may only be performed by qualified electricians. The regulations in the separate service

manual must be observed.