Operation and Control

Setting the relative humidity of the third ramp segment:

Turn the push-turn control to the right until the humidity

display flashes.

Hold down the SET key and set the desired humidity

setpoint to 50.0 % rh with the push-turn control.

%

rh

22. Setting the CO

2

content of the second ramp segment

(only for appliances with CO

2

supply, otherwise please

continue with point 23):

Turn the push-turn control to the right until the CO

2

display flashes.

Hold down the SET key and set the desired CO

2

setpoint

to 0.0% with the push-turn control.

CO

2

23. Setting the lighting of the second ramp segment (only

for models with lighting, otherwise please continue with

point 24):

Turn the push-turn control to the left until the lamp dl

display appears.

Hold down the SET key and set to on with the push-turn

control.

24. Setting the close statement of the third ramp segment:

Turn push-turn control to the right until a close state-

ment appears in the display, e.g. end, and press the SET

key briefly to confirm.

25. Leave programming write mode EDIT

Turn push-turn control to the right until EXIT appears in

EXIT

EXIT

the display, and press the SET key briefly to confirm.

26. Setting the temperature monitoring:

Turn the push-turn control to the right and adjust the

temperature monitoring (for more detailed information

on this, see page 53 ).

°C

MIN

AUTO

MAX

27. Activating the programme:

Turn the push-turn control to the right until the stop

icon flashes.

flashes.

Hold down the SET key and select the start icon with

with

the push-turn control. Release the SET key, and the pro-

gramme is activated.

5.10.8 Operation with PC/laptop (optional)

The cooled incubator can optionally be used, controlled and programmed with a PC/laptop.

For this purpose, it is equipped with a USB or ethernet interface on the rear side (see page 14

and 57 ).

The control of the appliance with the Memmert computer software “Celsius“ is

described in a separate manual.