Warning Messages and Malfunctions

6. Warning Messages and Malfunctions

6.1 Warning messages

The warning messages also set off an intermittent acoustic signal:

This can be temporarily switched off by pressing the SET key.

Error in the temperature control system (see also chapter "

Error in the temperature control system (see also chapter " Temperature monitoring

Temperature monitoring "

Temperature monitoring

Temperature monitoring

on page

on page 53 ): ):

53

53

Tb active - if the

Tb active -

Tb active -

temperature limiter is

triggered

Remedy: See page 54

hi-alarm - if

hi-alarm -

hi-alarm -

overtemperature

protection is triggered

Remedy: Check the setting of the

MAX temperature monitor.

if

undertemperature

protection is triggered

Remedy: Check the setting of the

MIN temperature monitor.

asf-alarm - if the

automatic monitoring

function is triggered

Remedy: Check the setting of the

ASF temperature monitor.

Error in the dehumidification system:

Error in the dehumidification system:

if the water supply is

defective

Remedy: Check that the water

supply is properly connected. Fill

water supply tank with distilled

water if it is empty.

If humidity exceeds the

setpoint for longer than

30 minutes.

Remedy: Open the door for

30 sec. and wait to see if the

controller steadily adjusts to

the setpoint. If the error occurs

again, contact the customer

service.

Error in the CO

22

system (only for models with CO

system (only for models with CO

22

supply)

supply)