3 replacing fluorescent tubes – Memmert ICH 256 Climate Chambers (Generation 2003) User Manual

Page 66

Advertising
background image

66

Maintenance and Servicing

8.3 Replacing fluorescent tubes

(only for models with interior lighting)

Caution:

Replace the fluorescent tubes only with fluorescent tubes of the same type; e.g. UV

light tubes only with UV light tubes, not daylight tubes. For information on the speci-

fications, please refer to chapter " Technical data " on page 17 .

Do not replace single tubes, but always the entire tube set (e.g. all UV light tubes

instead of just one). Replacing single tubes can influence the light output.

1. Loosen the connection of the illumination box in the interior. To do so, turn the bayonet

catch to the left and pull out the plug ( Fig. 34 ).

2. Pull the illumination box out of the appliance on both handles ( Fig. 35 ).

Fig. 34

Open the bayonet

catch and pull off the plug

Fig. 34

Fig. 34

Open the bayonet

Open the bayonet

Fig. 35

Pull out the illumination box

3. Turn the illumination box around and place it carefully on a smooth surface so that the

fluorescent tubes are on top ( Fig. 36 ).

4. Remove the retaining plate of the glass cover on one side. To do so, unlock the Allen

screws on the side, take off the retaining plate and pull out the glass cover ( Fig. 37 ).

Fig. 36

Turn around the illumination

box

Fig. 36

Fig. 36

Fig. 37

Remove Allen screws and retaining plate

and take out the glass cover.

Fig. 37

Fig. 37

Remove Allen screws and retaining plate

Remove Allen screws and retaining plate

Advertising