4 setting up – Memmert ICH 256 Climate Chambers (Generation 2003) User Manual

Page 22

Advertising
background image

22

Delivery, Transport and Setting Up

3.4 Setting up

The cooled incubator may only be placed on the ground, not on a table. When doing this,

ensure that the appliance is positioned exactly horizontally

.

The installation site must be level

and must be able to reliably carry the weight of the cooled incubator (see page 17 ). Do not

place the appliance on inflammable surfaces and do not stack.
The front swivel castors can be locked with a catch. To ensure stability, always turn the front

swivel castors to the front to lock them.
A power connection corresponding to the information on the nameplate (see page 17 ) must

be available at the installation site.
The distance between the wall and the rear of the cooled incubator must be at least 15 cm.

The clearance from the ceiling must not be less than 20 cm and the side clearance from the

wall must not be less than 8 cm ( Fig. 8 ). The ventilation slits of the cooling unit on the rear of

the appliance may not be adjusted.
Sufficient air circulation in the vicinity of the appliance must be guaranteed at all times.

≥ 15 cm

≥ 8 cm

≥ 8 cm

≥ 20 cm

Fig. 8

Minimum clearance from walls and ceiling