Operation and Control

The chamber must not be loaded too tightly, so that proper air circulation in the working

chamber is guaranteed. Do not place any of the chamber load on the floor, touching the side

walls or right below the ceiling (heating ribs) of the working chamber. To guarantee optimum

air circulation, push in the sliding grids so that the gaps between the door, sliding grid and

rear wall of the chamber are roughly the same size.

Fig. 14

Correct and incorrect chamber loading

5.6 Connecting gas supply

(only for models with CO

2

supply)

1. Check that the gas bottle is properly connected (see page 24 ).
2. Open the valve on the gas bottle.