Operation and Control

5.10.7 Settings example programme mode

On Monday at 8.00, the cooled incubator should heat up to 37 °C as quickly as possible with

a fan speed of 30 %, and reach a relative humidity of 70 % rh. For models with CO

2

supply,

the CO

2

content should be set to 5 % and for models with interior lighting, daylight should be

simulated.
Once the temperature and humidity have been reached, the cooled incubator should retain

the setpoint values at a fan speed of 50 % for 45 minutes. For models with lighting, the test

samples should be subject to UV light and daylight during that time.
Afterwards, the appliance should cool down to 20 °C and reach 50 % rh humidity within 1

hour at 30 % fan speed. For models with CO

2

supply, the CO

2

content should be adapted to

the natural content of the ambient air and for models with interior lighting, daylight should

be simulated. The overtemperature protection should trigger at 38.5 °C.

Mo 8.00 h

50 °C

37 °C

20 °C

50 %

30 %

80 %

0.45 h

1.00 h

t

0.01 h

70 % rh

50 % rh

10 %

15 %

5 %

Ramp 1

Ramp 2

Ramp 3

overtemperature limit

DL

UV

no

combination

possible

Fig. 20

Settings example programme mode