Warning Messages and Malfunctions

if the CO

2

supply is defec-

tive

Remedy: Set CO

2

setpoint to

0, check the stop valve and

connection of the gas bottle; if

the gas bottle is empty, replace

it and set the CO

2

setpoint to the

desired value.

if the CO

2

concentra-

tion exceeds the defined

setpoint by at least 1 % for

more than 3 minutes.

Remedy: Open the door for

30 sec. and wait to see if the

controller steadily adjusts to the

setpoint.

If the error occurs again, contact

the customer service.

6.2 System/appliance errors

Warning!

After removing covers, live parts may be exposed. You may

receive an electric shock if you touch these parts. Malfunctions

requiring intervention inside the appliance may only be rectified

by electricians. Please refer to the separate ICH cooled incubator

service manual for this.

Do not try to rectify the error yourself but contact an authorised customer service point for

MEMMERT appliances or contact the MEMMERT customer services department directly (see

page 2).
In case of enquiries, please always specify the model and appliance number from the

nameplate (see page 17 ).

Error

Possible cause

Remedy

does not work,

although the

cooled incubator

is switched on.

Power supply interrupted Check the power supply and fuse/circuit

breaker.

Appliance fuse or

miniature fuse or

controller defect

Contact the customer service and refer to the

service manual.

Appliance error

Power module defect

Appliance

cannot be

operated

Cooled incubator locked

with user-ID card

Unlock incubator with user-ID card (see page

59 ).

Push-turn control defect

Contact the customer service and refer to the

service manual.

No CO

2

display

shown in the

CO

2

module

Switched mode power

supply SP 200 defect

Contact the customer service and refer to the

service manual.