Advanced Functions

For each selected balance point ( Fig. 28 ), a positive or negative compensation correction value

can be set between -2.9 ºC and +2.9 ºC. There must be a difference of at least 10 ºC between

each of the individual balance points.

CAL. 1

+0,2 °C

CAL. 3

+0,3 °C

CAL. 2

-0,4 °C

Default

calibration

30°C

20°C

10°C

40°C

Fig. 28

Temperature calibration (example)

Setting:

Setting:

1. Choose SETUP and set the desired compensation temperature (see page 53 ) and the

accompanying compensation correction value to 0.0 °C.

2. Use a reference instrument to measure the deviation for the selected compensation

temperature.

3. Under SETUP, adjust the compensation correction value. If the measured reference

temperature is too low, the compensation correction value must be set with a negative

sign.

4. Perform a control measurement with the reference instrument.
5. If necessary, repeat the procedure for the other two balance points

Example: Temperature deviation at 35 °C is to be corrected.
1. Set compensation temperature CAL.2 under SETUP to 35.0 °C and set the corresponding

compensation correction value to 0.0 °C:

h

°C

°C

2. With a calibrated reference instrument and at a defined setpoint temperature of 35 °C in

normal operation, an actual temperature of 34.6 °C is measured.