Rockwell Automation 1785-LTx,D17856.2.1 Classic PLC-5 Programmable Controllers Users Manual User Manual

Page 119

Advertising
background image

Calculating Program Timing

Chapter 9

9-2

processor internal checks

updating the input image table with processor-resident I/O input status

updating processor-resident local I/O output modules with data from the

output image table

updating the input image table with remote I/O input status as contained

in the remote I/O buffer

updating the remote I/O buffer with output data from the output

image table

If no change in input status occurs and the processor continues to execute
the same logic instructions, the program scan cycle remains consistent (in
our example, at 25 ms). In real systems, however, the program scan cycle
fluctuates due to the following factors:

false logic executes faster than true logic
different instructions execute at different rates
different input states cause different sections of logic to be executed
interrupt programs affect program scan times

Effects of False versus True Logic on Scan Time
The rung below—which changes states from one program scan to the
next—changes your program scan time by about .25 ms.

MASKED MOVE

Source

Mask

Dest

N7:0

1100110011001100

N7:2

I:000

00

MVM

If I:000/00 is: Then the rung is:
On

True and the processor executes the maskedĆmove instruction. A maskedĆmove

instruction takes 258

µ

s to execute (see appendix A of the PLCĆ5 Programming

Software Instruction Set Reference, publication 6200Ć6.4.11).

Off

False and the processor scans the rung but does not execute it. It takes only 1.4

microseconds to just scan the rung.

Other instructions may have a greater or lesser effect.

Effects of Different Instructions on Logic Scan Time
Some instructions have a much greater effect on logic scan time than
others based on the time it takes to execute that instruction.

Program scan time is also affected by the basic construction of your ladder
rungs. The sizes of rungs and the number of branches in each can cause
the scan time to fluctuate greatly.

Advertising