Metrohm 915 KF Ti-Touch User Manual
Page 106

11.5 USB Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
92
■■■■■■■■
915 KF Ti-Touch
Rotation offset
The rotation offset is the offset from the center of the tower to the center
of the robotic arm. This value does not usually need to be changed. If a
Swing Head is to be mounted on the tower with a lateral offset, then this
value can be determined by the service technician when the rack is adjus-
ted.
Input range
–270.00 - 270.00 °
Default value
0.00 °
Swing direction
The swing direction of the robotic arm depends on its type. For a 2-tower
model, the robotic arm must be defined as right-swinging on Tower 1 and
as left-swinging on Tower 2.
Selection
+ | -
Default value
-
+
Left-swinging.
-
Right-swinging.
Properties – External position
Swing Head: Properties
▶ Edit
Properties which apply only to the selected external position can be
defined in the dialog Properties / Edit external position.
Angle
Swing angle for the selected external position.