23.6 Saving a determination

■■■■■■■■■■■■■■■■■■■■■■

210

■■■■■■■■

915 KF Ti-Touch

23.6

Saving a determination

Results dialog: More data

▶ Load/ Save ▶ Save

Determinations can be saved as follows:

■

Manually in this dialog.
The suggested default setting for file names is the first 16 characters of
Identification 1, followed by the determination time in the format
YYYYMMDD-hhmmss. Extending the file name with the determination
time has the advantage that the file names of all determinations are
unique.

■

Automatically at the end of the determination (see Chapter 16.5.6,
page 167)
.

The following data is stored for each determination:

■

Calculated results

■

Measuring point lists, equivalence points, curves and variables which
were generated during the determination

■

Determination properties

■

Statistics (if carried out)

■

Method with which the determination was carried out

■

Common variables (values and properties of all common variables at
the time of the determination)

■

Titrant data of the titrant used

23.7

Curves

Results dialog: Curve

The curve of the current determination is displayed in the Results / Curve
dialog. You can configure the curve display.

Advertising