Metrohm 915 KF Ti-Touch User Manual

Page 304

Advertising
background image

28.4 Calculations

■■■■■■■■■■■■■■■■■■■■■■

290

■■■■■■■■

915 KF Ti-Touch

NOTE

The titer is assigned to the titrant defined in the last titration command
before the calculation. Please note that the CALC command containing
the titer assignment is inserted after the titration command with which
the titer is determined.

Selection

Single value | Mean value | off

Default value

off

Single value
The result of the current determination is saved as titer.
Mean value
If statistic calculations are carried out for the result, the current mean
value of the determination series is saved.
off
The result is not saved as titer.

Save as common variable

on | off (Default value: off)

The calculated result can be saved as a method-independent variable, a
so-called common variable. The result is then also available in other meth-
ods for calculations. In all cases, only the single value will be saved, even if
the statistics function is enabled.

Variable

Selection of the common variable to which the result is being assigned.

Selection

CV01 … CV25

Display result

on | off (Default value: on)

If you deactivate this parameter, the result is neither displayed in the result
dialog nor printed in the result report. This can be advisable for intermedi-
ate results.

Save result in result table

on | off (Default value: off)

The calculated result can be saved in the result table. This may be advisa-
ble if e.g. the results of all determinations carried out on a particular day
are to be displayed clearly. A maximum of nine results from a determina-
tion can be saved in the result table.

Advertising