31.6 Result variables as parameter setting
■■■■■■■■■■■■■■■■■■■■■■
344
■■■■■■■■
915 KF Ti-Touch
31.6
Result variables as parameter setting
For most of the numerical parameters, the result of a calculation can also
be utilized in place of a number. This result must be calculated in a pre-
ceding calculation command (command CALC). A typical application is
the use of a relative start volume.
The following example is used to explain how this method is used to
define a sample size-dependent start volume for a titration. Proceed as
follows:
1 Create a new calculation
■
Insert the calculation command CALC in front of the titration
command.
■
Enter the calculation formula, e.g. R1 = C00*3.
2 Define a result variable as parameter setting
■
In the titration command, enter the variable R1 as the start vol-
ume.