Metrohm 915 KF Ti-Touch User Manual
Page 76

8.6 GLP test for exchange unit and dosing unit
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
915 KF Ti-Touch
GLP test interval
If you define a time interval for the GLP test, then the date in Next GLP
test will be tracked automatically.
Input range
1 - 999 days
Default value
999 days
Next GLP test
If you define a date for the next GLP test, then the GLP test interval will
be tracked automatically.
Format: YYYY:MM:DD
Action
Selection of the action which is carried out when the time interval has
expired.
Selection
Display message | Document message | Cancel
determination
Default value
Display message
For all three options it is documented in the deter-
mination data (see dialog More determination
data / Messages), that the time interval has been
expired.
Display message
A message is displayed. You can select whether you want to continue
with the determination or cancel the run.
Document message
In the determination data it will be documented, that the time interval
has been expired.
Cancel determination
The determination is stopped.