3 canceling determinations manually – Metrohm 915 KF Ti-Touch User Manual
Page 205

■■■■■■■■■■■■■■■■■■■■■■
21 Determination run
915 KF Ti-Touch
■■■■■■■■
191
21.3
Canceling determinations manually
You can cancel a determination at any time with the fixed key [
]. The
command being carried out at that moment is canceled and no further
commands are executed.
If you are working with the autostart function activated (see Chapter 17,
page 170), the entire series will be canceled.
NOTE
If you do not wish to cancel the entire determination, but only a single
command, use the function [Cancel command] in the live display. This
is particularly useful if you want to cancel a titration because the
sought-after equivalence point has already been determined. All subse-
quent commands, e.g. calculations and print report, will nevertheless
still be carried out.