Metrohm 915 KF Ti-Touch User Manual
Page 281

■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
915 KF Ti-Touch
■■■■■■■■
267
off
No drift correction takes place.
Drift value
This parameter can only be edited with Drift correction = manual.
Drift for manual drift correction.
Input range
0.0 - 99.9 µL/min
Default value
0.0 µL/min
Measured value display
on | off (Default value: off)
If this parameter is activated, the currently measured value is displayed
during the conditioning.
Conditioning options [Cond. options]
Cond. stop volume
Maximum permissible volume that can be dosed during conditioning.
Conditioning is stopped when the specified volume is dosed. If condition-
ing is continued by pressing [
] once again, then the titrant volume that
has already been dosed will not be taken into account; i.e. the dosing
starts again at zero. The stop volume should be adjusted to the size of the
titration cell in order to prevent any overflow.
Input range
0.00000 - 9999.99 mL
Default value
20.0000 mL
Selection
off
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Input range
1 - 999999 s
Selection
off
Default value
off
Delay reconditioning
on | off (Default value: off)
If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.